9 Mistakes To Avoid In Your SWAS Specs
Create perfect fossil power plant SWAS specs every time
The Sentry DCCP Reboiler Degassed Cation Conductivity Panel is designed to optimize the analytical results for degassed cation conductivity. First, the panel uses an oversized heater and baffled vapor discharge chamber, which assures complete boiling and provides a steam blanket over the sample to eliminate any sample contamination that might occur due to discharge piping backflow on the exhaust vent. Second, after the sample is degassed, it is cooled with an additional sample cooler so that chemistry deviations in analyzer temperature compensation curves are reduced.
The DCCP is a complete panel that measures specific cation and degassed conductivity or any individual measurement needed. The custom analysis panels include degassed conductivity, dissolved oxygen (DO) and corrosion products sampling. The panel is designed to work with a wide variety of conductivity analyzers and provides simple and compact cycle chemistry validation.
Module Type | Degassing |
---|---|
Max. Flow Rate (cc/min) | 200 |
Max Process Pressure | 100 psig (6.9 barg) |
Process Temperature | 130°F (54°C) |
Flow Rate | 100–200 cc/min |
Input Power | 110–120 Vac, 1 phase, 60 Hz or 220–240 Vac, 1 phase, 50 Hz |
Simplify the conditioning of steam, water, or non-hazardous process samples. Single conditioning modules provide signal conditioning for many types of signals. This improves the overall sampling process. There are fully integrated sample analysis packages for water, degassed cation conductivity, and steam available through Sentry Equipment. If you have any questions about the right single-line conditioning module for your desired application, please contact our team of engineers for more information. Sentry Equipment has industry-leading service and support for our clients through our technical support team.